腹腔镜维修常见品牌型号:
stryker史赛克:502-457-030 30°腹腔镜;502-457-010 0°腹腔镜;
olympus奥林巴斯:wa53005a 30°腹腔镜、a5294a 0°腹腔镜、a5295a 30°腹腔镜等;
storz史托斯:26003aa 0°腹腔镜、26003ba 30°腹腔镜、26003fa 45°腹腔镜、26046ba 30°小儿腹腔镜、26005aa 0°小儿腹腔镜等;
tiansong天松:e0001.1 0°腹腔镜、e0002.1 30°腹腔镜;
gimmi吉米:e.8293.11 30°腹腔镜;
hawk好克:th9333 30°腹腔镜;
wolf狼牌:8934.431 0°腹腔镜、8934.432 30°腹腔镜、8934.433 50°腹腔镜、8930.421 0°小儿腹腔镜、8930.422 30°小儿腹腔镜、8930.451 0°小儿腹腔镜等;
广州奥得富***——维修腹腔镜
另外,广州奥得富***还提供以下相关内窥镜周边维修服务
硬镜维修:椎间孔镜维修、输尿管镜维修、经皮镜维修、腹腔镜维修、膀胱镜维修、宫腔镜维修、关节镜维修、鼻窦镜维修、电切镜维修、耳镜维修等。
![](http://img3.dns4.cn/pic/280403/p6/20190522150315_9855_zs_sy.jpg)
广州奥得富***是一家***提供腹腔镜硬管性内窥镜第三方服务商。***提供STORZ 26011BA 30° 小儿腹腔镜维修,德国STORZ硬管镜维修。德国STORZ(史托斯)内窥镜包含有:史托斯腹腔镜、STORZ宫腔镜、史托斯输尿管硬镜、STORZ经皮镜、史托斯鼻窦镜等硬/软性内窥镜维修。图像闪烁,电子腹腔镜维修,ccd老化,电子腹腔镜维修ccd驱动板损坏,偶有雪花状。要了解更多STORZ内窥镜维修、史托斯腹腔镜等维修资讯,欢迎咨询!
STORZ腹腔镜维修-斯托斯硬镜维修维修硬镜/宫腔镜维修/电子腹腔镜维修
设备名称:STORZ 26011BA 小儿腹腔镜
故障现象:没有图像,内部光学部件损坏,镜管弯曲
检测结果:镜头内部镜片断裂,灰尘较大,图像模糊
解决方案:更换内部光学部件,***维修电子腹腔镜,处理镜管问题,重新调整图
奥得富***
OLYMPUS电子腹腔镜的故障作简单介绍。
(一) 导光管故障1、导光管老化,脱胶。2、导光管折断。 维修:更换导光管。
(二) 图像故障:1、图像失真,驱动板坏、线缆接触不良。2、无图像,A50001A电子腹腔镜维修,线缆断或CCD坏。
(三) 导光束故障;1、无亮度,光束折断。2、亮度偏暗,光束折断、导光头进水或光束烧焦。此种故障维修方案,更换导光束。
(四) 按键故障。1、按键失灵或时有时无,更换按键。2、按键线缆老化或断开,电子腹腔镜维修,更换按键控制线缆。
冲洗、干燥 :
内镜从消毒槽取出前,清洗消毒人员应当更换手套,用无菌纱布或洁净压缩空气等方法进行、干燥。化学***浸泡灭菌的内镜,使用前必须用无菌水彻底冲洗,去除残留***。
储存 :
将消毒或灭菌的内镜存放在灭菌后的密闭器械盒内。