17β-Estradiol (E2) ELISA Kit 17β雌二醇E2试剂盒
- E2 monoclonal antibody binds exclusively with E2 and does not show cross-reaction with other chemicals of similar structures. A monoclonal antibody is uniform in quality, generating very little lot-to-lot variation.
- The quantitative analysis ranges from 0.05μg/L to 1μg/L (ppb).
- The ELISA measurement is highly reproducible; the coefficient of variation (CV) is mostly under 10%.
- The assay requires less amount of harmful solvent than instrument analyses.
- With ease of handling, the total time for measurement is only 2.5 hours.
- The kit, a 96-well microplate format, enables simultaneous measurement of multiple samples at more reasonable cost.
Chemical Structure
|
Standard Curves
|
Comparison with LC/MS/MS
Averaged value of E2 (ng/L)
|
|||||||
Sample
|
Na)
|
LC-MS/MS
|
E2-Ecologienab)
|
E2-ADc)
|
E2-CCc)
|
E2-NGc)
|
E2-RBc)
|
Primary effluent
|
10
|
8.4
|
12.5
|
21.1
|
81.4
|
97.6
|
202.2
|
Aeration tank
|
13
|
1.9
|
2.1
|
4.4
|
9.4
|
7.8
|
31.6
|
Secondary effluent
|
5
|
1.2
|
2.6
|
5.3
|
15.1
|
18.8
|
75.9
|
|
|
a)
|
N=number of samples
|
b)
|
Ecologiena E2 kit
|
c)
|
Other kits on the market (for biological samples)
|
Cross-reactivity Pattern
Estrogens
|
|||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||
|
|
|||||||||||||||||||
Other hormones
|
|||||||||||||||||||
|
|||||||||||||||||||
|
Kit Format
Kit Format
|
Comment
|
|
Microplate
|
1 plate (96 wells)& reagents
|
Needs a microplate reader (450nm)
For multiple and simultaneous measurement |